![HP 8562E Service Manual Download Page 399](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511399.webp)
1. If the YTO PLL is unlocked, error code 301 should be displayed. Place the spectrum
analyzer in ZERO SPAN. Figure 11-7 illustrates the simplified YTO PLL.
2. Move the jumper on
to connect pins 2 and 3 (TEST position). Refer to
Figure
for the location of
Error code 301 should no longer be displayed.
(The YTO PLL feedback path is now open and the YTO error voltage is forced to zero.)
3. On the spectrum analyzer, press (CAL),
MORE
2,
, and LO
.
The displayed LO FREQ is the desired YTO frequency calculated. Record the calculated
frequency of the YTO below:
YTO Frequency (calculated) =
MHz
4. Measure the YTO frequency at the front-panel First LO OUTPUT jack and record below:
YTO Frequency (measured) =
MHz
1
I
I
I
Figure 1
Troubleshooting an Unlocked YTO PLL
Synthesizer Section
1 l-25