![HP 8562E Service Manual Download Page 290](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511290.webp)
2. Disconnect W26 (coax 2) from
and connect the output of a function generator to
3. Set the function generator to the following settings:
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..Sinewave
Amplitude
. . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 V pk-to-pk
DC Offset
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4. Set the spectrum analyzer sweep time to 50 ms.
5. Adjust the function generator amplitude and offset until the sine wave fills the entire
graticule area.
6. Measure and note the function generator peak-to-peak voltage using an oscilloscope.
V
7. Set the spectrum analyzer to 2 dB/div.
8. Readjust the function generator amplitude and offset until the sine wave again fills the
entire graticule area.
9. Measure the function generator peak-to-peak voltage and dc offset.
V
10. The ratio of voltage recorded in step 6 to the voltage recorded in step 9 should be 5
If the ratio is not 5, troubleshoot the A3 Interface assembly.
11. Reconnect W26 to
Refer to function block U of A3 Interface Assembly Schematic Diagram in the
HP 8560 E-Series Spectrum Analyzer Component Level Information.
1. Press
and set the spectrum analyzer controls as follows:
Center frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2. Press
(CAL), and IF
OFF. Connect the CAL OUTPUT to the INPUT
connector.
3. Check for a TTL high on
pin 2 and a TTL low on
pin 10. Set the spectrum
analyzer to 2 dB/div and check for a TTL high on
pin 10 and a TTL low on
pin 2.
4. If the logic levels on
are incorrect, check the LLOG-STB signal
follows:
a. Monitor
pin 9 with an oscilloscope or logic probe. Check that a 1
microsecond, low-going pulse is present when switching between 10 dB/div and
2 dB/div.
b. Check the inputs to
(pins 3 and 11) while switching between 10 dB/div and
2 dB/div.
8-16
Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......