![HP 8562E Service Manual Download Page 159](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511159.webp)
A10 YIG-Tuned Filter/Mixer
Do NOT remove the brackets from the A10 assembly. If these brackets are
removed and reinstalled, the performance of A10 will be altered. A new or
rebuilt A10 assembly includes new mounting brackets already attached to it.
1.
2.
3.
4.
Disconnect W16 ribbon cable from the A10 assembly.
Use a
wrench to remove W43, W44, and W46 from
J2, and J4,
respectively.
Remove
( as a unit) and disconnect W48 from
Remove four screws (1) securing A10 to the center deck. These screws are located on the
top-side of the center deck as illustrated in Figure 4-16.
( 4 P I
( 3 P I
Figure 4-16. A10 Mounting Screws
1. Orient the A10 assembly for the proper cable connections.
2. Connect
and 46 to
J2, and J4, respectively.
3. Connect
(gray cable) to
4. Connect W16 ribbon cable to the A10 assembly.
5. Install
and ensure that all cable connections are tight. Torque them to 1
Ncm (10 in-lb).
6. Secure A10 to the spectrum analyzer center deck using four screws.
13
4-34 Assembly Replacement
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......