5. Compare the signals at the following test points with those illustrated in Figure 10-2.
x
pin 14
Y
pin 3
Z OUT:
pin 3
BLANKING:
pin 15
Waveforms displayed on an analog scope may show considerably more spikes.
This is normal and is due to the wider displayed bandwidth.
6. Troubleshoot the circuits associated with any bad waveforms.
3 . 0 0
0 . 0 0 v
1
m s / d i v
0 . 0 0 0
x
Y
z
B L A N K I N G
S K 1 9 1
Figure 10-2. Line Generator Output Waveforms
Refer to function block J of A2 Controller Schematic Diagram (sheet 1 of 4) in the
HP 8560 E-Series
Spectrum
Analyzer
Component Level Information.
1. Using an oscilloscope, check for blanking pulses at
pin 15.
pin 6 should be
at a TTL high. Blanking pulses turn the CRT beam off during the sample time of the line
generators and when moving the CRT beam to a new position for drawing the next vector.
2. Set an oscilloscope to the following settings:
Amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Amplitude offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
V
Sweep time
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Triggering
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . External
3. Externally trigger the oscilloscope off the signal at
pin 8 (LBRIGHT).
Controller Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......