Procedure 12.
Battery
Procedure 12.
Battery
Battery
contains lithium polycarbon monofluoride. Do not incinerate or
puncture this battery. Dispose of discharged battery in a safe manner.
To avoid loss of the calibration constants stored on the A2 controller assembly,
connect the spectrum analyzer to the main power source and turn on before
removing the battery.
The battery used in this instrument is designed to last several years. An output voltage of
V is maintained for most of its useful life. Once this voltage drops to
V, its life and
use are limited and the output voltage will deteriorate quickly. When the instrument is turned
off, stored states and traces will only be retained for a short time and may be lost. Refer to
“State- and Trace-Storage Problems” in chapter 10, “Controller Section,” in this manual. The
battery should be replaced if its voltage is
V or less.
1. Remove any option module attached to the rear panel.
2. Locate the battery assembly cover on the spectrum analyzer rear panel. Use a screwdriver
to remove the two
screws securing the cover to the spectrum analyzer.
3. Remove the old battery and replace it with the new one, ensuring proper polarity.
4. Measure the voltage across the new battery. Nominal new battery voltage is approximately
V. If this is not the case, check the battery cable and A2 controller assembly.
5. Secure the battery assembly into the spectrum analyzer.
Assembly Replacement 4-43
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......