![HP 8562E Service Manual Download Page 446](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511446.webp)
4. With the SIG ID Oscillator on, measure the frequency at
with a frequency
counter and an active probe.
5. On the HP 85623, press
until
pin 13 is at TTL low. Diodes CR603 and
CR605 should be forward biased and CR604 should be reverse biased (approximately 6 Vdc
reverse bias). Diodes CR501 and CR502 should be forward-biased, disabling the 3rd LO
Driver Amplifier.
6. The voltage at the
node should measure approximately -5 Vdc, biasing
Q604
on.
7. If oscillator bias voltages are correct, place a 100 resistor across SAWR
input
and output. If the SAWR has failed, this will provide the equivalent loss of a correctly
functioning SAWR, and the circuit will begin to oscillate.
The spectrum analyzer 10 MHz reference consists of 10 MHz OCXO (Option 103: TCXO)
with associated TTL level generator and distribution circuitry. The OCXO (or TCXO) and
TTL level generator are turned off when an external 10 MHz reference is used. Also, with
the analyzer set to EXTernal frequency reference,
output (low) forces the output of
to stay high. This allows
to control the outputs of
and
In
frequency reference,
controls the outputs of these three NAND gates, and
the output of
is held high.
Check the 10 MHz reference by performing the following steps:
1. Set the HP 85623 10 MHz reference to internal by pressing
REAR
, and
MHz INT.
2. Use a spectrum analyzer to confirm the presence of a 10 MHz signal at the following test
points:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3. Check for a 1.3 Vp-p waveform at
using an oscilloscope (see Figure 12-6).
4. Check that the signal at
is 10 MHz
Hz (with Option 103 TCXO reference)
using a frequency counter. If necessary, perform the appropriate 10 MHz reference
adjustment.
5. If there is no problem with
10 MHz reference operation, check EXTernal 10 MHz
reference operation as follows:
6. Set the HP 85623 10 MHz reference to external by pressing
EXT.
7. Connect a 10 MHz, -2
signal to the rear-panel 10 MHz REF IN/OUT connector.
8. Check the signals at
and
according to the
procedure in steps 2 through 4.
9. If the signals are correct in EXTernal operation, but not in
operation, the
problem lies in A21 OCXO (or Option 103 TCXO), its voltage reference, or the TTL level
generator. Check these areas as follows:
a. On the spectrum analyzer, press 10 MHz INT .
12-20 RF Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......