Procedure 5. A2, A3, A4, and A5 Assemblies
Procedure 5.
A2, A3, A4, and A5 Assemblies
1. Remove the spectrum analyzer cover.
2. Place the spectrum analyzer on its right side frame.
3. Remove the eight screws holding the A2, A3, A4, and A5 assemblies to the top of the
spectrum analyzer. These screws are labeled
and (4) in Figure 4-7. They are also
labeled on the back of the A2 board assembly.
4. Remove ribbon cable W4 from
See Figure 4-7.
Do not fold the board assemblies out of the spectrum analyzer one at a time.
Always fold the A2 and A3 assemblies as a unit and the A4 and A5 assemblies
as a unit. Folding out one assembly at a time binds the hinges attaching the
assemblies and may damage an assembly and hinge.
5. The board assemblies are attached to the right side frame of the spectrum analyzer with
two hinges. Fold both the A2 and A3 assemblies out of the spectrum analyzer as a unit.
6. Fold both the A4 and A5 assemblies out of the spectrum analyzer as a unit.
7. Remove the cables from the assembly being removed, as illustrated in Figure 4-8.
8. Remove the two screws that attach the assembly being removed to its two mounting hinges.
Do not torque shield TORX screws to more than 8 inch-pounds. Applying
excessive torque will cause the screws to stretch.
Assembly Replacement
15
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......