17. Tune the source 1
above the fractional N frequency. The voltage measured on the
DVM should be approximately -12 Vdc.
18. If the DVM reading does not change, the Al4 frequency control assembly is defective.
Reconnect W32 to
Replace the jumper on
to the NORM position.
SYNTHESIZED
SWEEPER
DIGITAL VOLTMETER
TEST CABLE
P I N 3
P I N 1
Figure 11-3. YTO Loop Test Setup
Check Al5 RF assembly (steps 19-25)
19. Disconnect W34 from
and disconnect W32 from
20. Connect a frequency counter to
Connect a high-frequency test cable from an
HP
synthesized sweeper to
See Figure 11-4.
21. Connect a BNC cable from the spectrum analyzer 10 MHz REF IN/OUT to the
HP
FREQUENCY STANDARD EXT input.
22. Set the HP
to the following settings:
Frequency standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EXT
Powerlevel
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -5dBm
23. Set the spectrum analyzer to the following settings:
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Single
24. Set the spectrum analyzer and HP
frequencies to the combinations listed in
Table 11-5 and press
on the spectrum analyzer.
Synthesizer Section
1
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......