18. 600 MHz Reference Adjustment
Al5 RF assembly
There is no related performance test for this adjustment.
The 100 MHz VCXO and the tripler are adjusted for a maximum signal level at 600 MHz.
A spectrum analyzer is used to monitor the amplitude of the 600 MHz signal while performing
these adjustments.
Spectrum analyzer
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
1. Press
to turn the spectrum analyzer off, disconnect the power cord, and remove the
spectrum analyzer cover. Fold down the Al5 RF and Al4 frequency control assemblies.
Prop up the Al4 frequency control assembly.
2. Disconnect W33, gray/brown coax cable, from
3. Connect the signal at
to the input of the HP
spectrum analyzer.
4. Reconnect the power cord and press
to turn the spectrum analyzer on.
5. Set the center frequency of the HP
to 600 MHz and set the frequency span and
resolution bandwidth of the HP
for the best display of the 600 MHz signal.
6. Set the peak of the 600 MHz signal near the top graticule line on the HP
display
and set to 1
per division.
7. Adjust Al5
VCXO Adjust, for maximum amplitude.
8. Adjust Al5 C751 Tripler Adjust, for maximum amplitude. The level, after proper
adjustment, should be between -3 and
(typically 0 to
9. Reconnect W33 to
2-66 Adjustment Procedures
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......