![HP 8562E Service Manual Download Page 442](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511442.webp)
Al5 RF Assembly
The block diagrams for the Al4 and Al5 assemblies are located in Chapter 11,
“Synthesizer Section.”
1. Perform the “IF Input Amplitude Accuracy” performance test in the
8560
E-Series
Spectrum Analyzer Calibration Guide.
This exercises most of the third converter.
2. If the performance test fails, perform the “External Mixer Amplitude Adjustment” in
Chapter 2, “Adjustment Procedures.”
3. If adjustment cannot be made, disconnect W35 (coax 92) from
4. On the HP 85623, press
and set the controls to the following settings:
Center frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. Connect a signal generator to
6. Set the signal generator to the following settings:
Frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CW
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -28dBm
7. If a flat line is displayed within 2
of the reference level and the performance test passed,
troubleshoot microcircuits A7, A8,
and A13.
8. If a flat line is displayed within 2
of the reference level and the performance test failed,
troubleshoot the Al5 RF assembly.
9. Disconnect the signal generator from
and reconnect W35.
1. Connect the HP 85623 CAL OUTPUT to the INPUT
connector.
2. Set the HP 85623 to the following settings:
Center frequency
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Span
0 Hz
Input attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.
Press
(CAL),
AD3 OFF.
4. Disconnect W29 (coax 7) from
5. Connect a test cable from
to the input of another spectrum analyzer.
6. Tune the other spectrum analyzer to 10.7 MHz. The signal displayed on the other
spectrum analyzer should be approximately -15
7. Remove the test cable from
and reconnect W29 to
RF Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......