900 TG UNLVL
901
902 BAD NORM
903 A DLMT
904 B DLMT
905 EXT REF
906
907 DO IF CAL
908
909
910
These error codes indicate user-generated errors.
Tracking generator output is unleveled.
Tracking generator output unleveled because START FREQ is set
below tracking generator frequency limit (300
The state of the stored trace does not match the current state of the
spectrum analyzer.
Unnormalized trace A is off-screen with trace math or normalization
on.
Calibration trace (trace B) is off-screen with trace math or
normalization on.
Unable to lock cal oscillator when set to external frequency reference.
Check that the external 10 MHz reference is within tolerance.
The oven-controlled crystal oscillator (OCXO) oven is cold.
Unit is still performing IF calibrations, or is in need of IF calibrations
which were not yet done due to an
condition, since an
error is indicative of a bandwidth 1
not getting
calibrated.
Channel bandwidth is too wide, compared to the channel spacing, for
a meaningful adjacent channel power computation.
The frequency span is too small to obtain a valid adjacent channel
power (ACP) measurement.
The frequency span is too wide, compared to the channel bandwidth,
to obtain a valid adjacent channel power (ACP) measurement.
General Troubleshooting 7-41
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......