![HP 8562E Service Manual Download Page 146](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511146.webp)
Procedure 6. A6 Power Supply Assembly
Procedure 6.
A6 Power Supply Assembly
The A6 power supply and
high voltage assemblies contain lethal voltages
with lethal currents in all areas. Use extreme care when servicing these
assemblies. Always disconnect the power cord from the instrument before
beginning this replacement procedure. Failure to follow this precaution will
present a shock hazard which may result in personal injury.
1. Disconnect the power cord from the spectrum analyzer.
2. Remove the spectrum analyzer cover assembly. Refer to “Procedure 1. Spectrum Analyzer
Cover
3. Fold out the A2, A3, A4, and A5 assemblies as described in “Procedure 5. A2, A3, A4,
and A5 Assemblies Removal,” steps 2 through 6.
4. Place the spectrum analyzer top-side-up on the work bench with A2, A3, A4, and A5
folded out to the right.
The voltage potential at
is
Disconnect at the CRT with caution!
Failure to properly discharge
may result in
severe electrical shock
to
personnel and damage to the instrument.
5. Connect the spectrum analyzer line-power cord to provide proper grounding while
discharging the
post-accelerator cable. Make sure that the spectrum analyzer
line-power switch is in the off position.
6. Connect a high voltage probe
such as the HP 34111A to a voltmeter with a
10
input.
7. Connect the clip lead of the probe (ground) to the chassis of the spectrum analyzer.
8. Slip the tip of the high voltage probe under the rubber shroud of the
post-accelerator cable to obtain a reading on the voltmeter. See Figure 4-2.
9. Keep the high voltage probe on the post-accelerator connector until the voltage has
dropped to a voltmeter reading of less than 5
(1ess than 5 V at the connector). This
normally takes about 30 seconds.
10. Disconnect the line-power cord from the spectrum analyzer.
11. Using a small screwdriver with the shank in contact with the CRT shield assembly, slip
the tip of the screwdriver under the rubber shroud of the
post-accelerator cable
and short the cable to ground on the CRT shield assembly. See Figure 4-2.
12. Remove the three screws securing the power supply shield to the power supply and remove
the shield.
13. Remove the three screws securing the
high voltage assembly to the A6 power supply
assembly.
14. Disconnect ribbon cable
from
and lift the
assembly out of the way.
See Figure 11.
Assembly Replacement 4-21
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......