Troubleshooting Using the TAM
Refer to Chapter 7, “General Troubleshooting,” for information on enabling the TAM for use
with the HP 85623 Spectrum Analyzer.
When using Automatic Fault Isolation, the TAM indicates suspected circuits that need to be
manually checked. Use Table 8-2 to locate the manual procedure.
Table 8-3 lists assembly test connectors associated with each Manual Probe Troubleshooting
test. Figure 8-l illustrates the location of the A3 test connectors.
A 3
I N T E R F A C E
C O N N E C T O R )
SK169
Figure 6-1. A3 Test Connectors
Analog data bus errors that occur during Automatic Fault Isolation result from either a
shorted W2 control cable or faulty A3 assembly. Perform the following steps to determine the
cause of the error.
1. Disconnect W2 from
and repeat the Automatic Fault Isolation procedure.
2. If the analog data bus error is still present, troubleshoot the A3 Interface assembly. If the
error disappears, look for a short on W2 or another assembly connecting to it.
3. To isolate a short on W2, reconnect W2 to
and disconnect W2 from all other
assemblies.
4. Repeat the Automatic Fault Isolation routine.
5. If the analog data bus error is still present, W2 is shorted. If the error disappears,
reconnect the other assemblies one at a time and repeat the procedure. Once the faulty
assembly is reconnected to W2, the error should reappear.
6-4
Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......