If the spectrum analyzer has excessive noise at
offset, measure the noise with center
frequencies of 100 MHz and 2.5
If the measurements are equal, suspect the fractional N circuitry and the YTO loop circuitry
on the Al4 frequency control assembly.
If the measurements differ by 2
to 5
with the 2.5
measurement at a higher noise
level, suspect the offset lock loop circuitry.
Check the noise on the 5 V regulators on A14, particularly the regulator in the reference
divider circuitry
Refer to function block AN on the Al4 frequency control assembly
schematic.
The noise level of the voltage regulator should be
The typical noise level is 40
RMS between 10 Hz and 100
A coaxial probe with very little unshielded tip area should be used to avoid picking up
radiated 60 Hz. Check that your measurement is valid by probing ground on the circuit
and verifying that the measured value is well under the 1
threshold that indicates a
defective regulator.
There can also be phase noise problems if the loop gain is incorrect. See function block AP for
loop gain troubleshooting information.
11-44 Synthesizer Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......