16. Use a DVM to monitor the voltage at
pin 4. The voltage should measure
approximately -8.45 Vdc.
17. Set the spectrum analyzer sweep time to 100 ms. The voltage at J18 pin 4 should measure
approximately -4.21 Vdc.
18. Set the spectrum analyzer sweep time to 200 ms. The voltage at J18 pin 4 should measure
approximately -2.1 Vdc.
19. The analog switches and comparators should be set as listed in Table 11-16.
20. Check that
opens when the sweep time is set to 2 seconds.
21. Check that
and
close when the sweep time is set to 20 seconds.
Sweep Time
200 ms
2 s
20 s
Table 11-16. Settings for Switches and Comparators
Switch
Switch
Switch
Comparator
Comparator
Pin 1
Pin 7
Open
Open
Closed
High
High
Open
Open
Open
High
High
Closed
Closed
Open
High
High
During multiband sweeps, the sweep ramp at
pin 15 should go from 0 V to
V
for each band or portions of a band covered. See function block A of Al4 frequency control
schematic in the
HP 8560 E-Series Spectrum Analyzers Component Level Information
binder.
However, the scan ramp at
pin 1 is scaled according to the percentage of the total
span that the band is covering. See function block B of Al4 schematic. Also, the sum of the
individual ramps is 10 V. Figure 11-9 illustrates both sweep and the scan ramp for a 0
to
13.2
span with instrument preset conditions.
Synthesizer Section 11-41
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......