8. LO Distribution Amplifier Adjustment
15. On
the HP 85623, press
MORE
2, SERVICE CAL DATA, LO LEVELS,
and
INT
LEVEL.
The message
DRIVE FOR
1
will be displayed.
16. Note the
INT Sense” voltage printed on the A7 LO distribution amp label. Use the
knob or keypad, and press enter, to change the displayed DAC value so the DVM reading
is 6
more negative than the label voltage. For example, if the
INT Sense” voltage
is -170
change the displayed DAC value so the DVM reading is -176
17. Record the DAC value:
DAC value for 4.60
18. To set the band 2 sense voltage, set the HP 85623 center frequency to 9.46
19.
On the HP 85623, press
(CAL),
OF 2, SERVICE CAL DATA,
LO LEVELS, and
INT LO LEVEL.
The message
DRIVE FOR
2
will be displayed.
20. Note the “B2 INT Sense” voltage printed on the A7 LO distribution amp label. Use the
knob or keypad, and press enter, to change the displayed DAC value so the DVM reading
is 6
more negative than the label voltage. For example, if the “B2 INT Sense” voltage
is 170
change the displayed DAC value so the DVM reading is 176
21. Set the “Sense EXT” value by pressing
EXT
LEVEL .
22. Use the knob or keypad to enter the DAC value for 1.45
from the band 0 sense
voltage adjustment above.
23. Save the adjustment values by pressing
MENU, STORE DATA,
and YES.
24. Move the jumper on
from WR
back to the WR PROT position.
2-42 Adjustment Procedures
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......