![HP 8562E Service Manual Download Page 386](http://html.mh-extra.com/html/hp/8562e/8562e_service-manual_161511386.webp)
BNC CABLE
FREQUENCY COUNTER
TEST CABLE
A 10
Figure 1 l-4. Sampler and Sampling Oscillator Test Setup
25. At each combination, the frequency counter should measure a sampler IF as shown in
Table 11-5. (The sampling oscillator of the offset PLL tunes to the frequencies listed in
the table.) If the frequency counter does not read the indicated sampler IF
suspect the Al5 RF assembly.
26. Reconnect W34 to
and W32 to
27. The First LO OUTPUT, located on the front panel, must be terminated in 50 ohms. If
the YTO unlocks only with certain center frequency and span combinations, check that
the termination is in place.
28. Set the spectrum analyzer
and (SPAN) to generate the unlock conditions.
29. On the spectrum analyzer press
30. Move jumper
to the TEST position.
31. Disconnect W34 from
and measure the power of the signal at the end of W34.
32. If the power is less than -6.5
suspect W34, A7 SLODA, or All YTO.
33. Move jumper
to the NORM position.
1 I-12 Synthesizer Section
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......