10. Calibrator Amplitude Adjustment
10. Calibrator Amplitude Adjustment
Al5 RF assembly
Calibrator Amplitude and Frequency Accuracy
The CAL OUTPUT amplitude is adjusted for -10.00
measured directly at the front
panel CAL OUTPUT connector.
Figure 2-18. Calibrator Amplitude Adjustment Setup
P C
BOARD
PROP
Measuring receiver
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8902A
Power sensor
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A
Adapters
Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1250-1477
S K 1 1 3
The spectrum analyzer should be allowed to warm up for at least 30 minutes
before performing this adjustment.
1. Place the spectrum analyzer in the service position shown in Figure 2-18. Prop the Al4
frequency control board assembly in the service position.
2. Zero and calibrate the HP
8482A combination in log display mode. Enter the
power sensor 300 MHz cal factor into the HP 8902A.
3. Connect the HP 8482A through an adapter directly to the spectrum analyzer CAL
OUTPUT connector.
4. Adjust
CAL AMPTD for a -10.00
reading on the HP 8902A display.
Adjustment Procedures 2-47
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......