The following measurements should be made with a high-voltage probe, such
as
the HP 34111A. When using the high-voltage probe, connect the ground
lead securely to the spectrum analyzer chassis.
7. Carefully measure the grid voltage at
pin 6, and the cathode voltage at
pin
4. The display will work with a cathode voltage of -2450 V
V, provided the grid
voltage
pin 6) is 30 to 100 V more negative than the cathode.
CUTOFF,
should be able to adjust the voltage difference over a 60 V range to account for tube
variations, and achieve proper intensity.
8. If the grid and cathode voltages are correct, turn off the spectrum analyzer and check
CR10 with an ohmmeter. If CR10 is good, suspect the
CRT.
9. If the grid and cathode voltages are too low, turn off the power and disconnect W8 from
the base of
CRT, and recheck the grid and cathode voltages.
10. If the grid and cathode voltages are still too low, refer to “CRT Supply” in this chapter
and the “High-Voltage Power Supply Adjustment” procedure in Chapter 2, “Adjustment
Procedures.”
11. If voltages are correct when the tube is disconnected, the CRT is probably defective.
The pins on the
CRT bend easily. Be careful not to bend pins when
connecting W8 to
Display/Power Supply Section
13-l 1
Summary of Contents for 8562E
Page 21: ...A l HP 85623 Interconnect Block Diagram A 5 Contents 14 ...
Page 205: ...8560E Series sl131e FIGURE 5 3 PARTS IDENTIFICATION MAIN CHASSIS v ...
Page 219: ...FL 4 Bl Figure 6 9 Rear View BTl A20 SK162 Major Assembly and Cable Locations 6 l1 ...
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ......
Page 314: ......
Page 355: ......
Page 356: ......
Page 357: ......
Page 374: ......
Page 425: ......
Page 426: ......
Page 450: ......
Page 469: ......
Page 470: ......
Page 474: ......