5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:LEVel
Sets the test signal level.
:LEVel?
Queries the test signal level.
Syntax
:LEVel
<data>
<data>
V/CV/CC (character data)
Function
Sets the test signal level to one of the followings.
V Open circuit voltage level
CV Constant voltage level
CC Constant current level
Example
Transmission
:LEVel CV
The test signal level is set to constant voltage.
Error
If <data> is other than numerical value described above, an execution error
occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:LEVel?
Function
Returns the test signal level as character data.
V Open circuit voltage level
CV Constant voltage level
CC Constant current level
Example
Response
If headers are on :
LEVEL CV
If headers are off
CV
The test signal level has been set to constant voltage.
Error
If the response message is longer than 300 bytes, a query error occurs.