46
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:CORRection:DATA?
Queries the open circuit and short circuit compensation values.
Syntax
:CORRection:DATA?
Function
Returns the open and short circuits compensation values at the currently test
frequency as <residual impedance of short circuit compensation>, <phase angle of
short circuit compensation>, <residual impedance of open circuit compensation>,
<phase angle of open circuit compensation> in order.
Residual impedance Numerical value in NR3 format or OFF (character data)
Phase angle Numerical value in NR2 format or OFF (character data)
When the compensation setting is OFF, or when the set test frequency of the
compensation differs from the current test frequency, returns the character data
"OFF."
Example
Response
If headers are on
:CORRECTION:DATA OFF,OFF,247.45E+06,-21.58
If headers are off
OFF,OFF,247.45E+06,-21.58
The short circuit compensation for the current test frequency is set to OFF, and
open circuit compensation is 247.45 M
Ω
, -21.58.
Error
If the response message is longer than 300 bytes, a query error is generated.