44
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:COMParator:SLIMit:MODE
Sets the second parameter setting mode for the comparator.
:COMParator:SLIMit:MODE?
Queries the setting mode of the second parameter for the
comparator.
Syntax
:COMParator:SLIMit:MODE
<data>
<data>
ABSolute/PERcent/DEViation (character data)
Function
Sets the second parameter setting mode for the comparator function.
ABSolute Absolute value setting mode (ABS)
PERcent Percentage value setting mode (%)
DEViation Deviation percentage setting mode (
∆
%)
Example
Transmission
:COMParator:SLIMit:MODE PERcent
The percentage setting mode is selected.
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:COMParator:SLIMit:MODE?
Function
Returns the current setting mode for the second parameter for the comparator
function as character data.
ABSOLUTE Absolute value setting mode (ABS) is selected.
PERCENT Percentage setting mode (%) is selected.
DEVIATION Deviation percentage setting mode (
∆
%) is selected.
Example
Response
If headers are on
:COMPARATOR:SLIMIT:MODE PERCENT
If headers are off
PERCENT
Error
If the response message is longer than 300 bytes, a query error is generated.