5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:PARameter1
(2, 3, or 4)
Sets the displayed parameters.
:PARameter1
(2, 3, or 4)
?
Queries the displayed parameters.
Syntax
:PARameter
1 (2, 3, or 4) <data>
<data>
Z Impedance
Y Admittance
PHASe Phase angle
CS Series equivalent static capacitance
CP Parallel equivalent static capacitance
D Loss coefficient
LS Series equivalent inductance
LP Parallel equivalent inductance
Q Q factor
RS Series equivalent resistance
G Conductance
RP Parallel equivalent resistance
X Reactance
B Susceptance
OFF
Function
Sets the displayed parameters.
Example
Transmission
:PARameter1 Z;:PARameter3 PHASe
The first parameter is set to impedance, and the third parameter is set to phase
angle.
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:PARameter
1 (2, 3, or 4)
?
Function
Returns the displayed parameters as character data.
Z, Y, PHASE, CS, CP, D, LS, LP, Q, RS, G, RP, X, B, OFF
Example
Response
If headers are on
:PARAMETER2 PHASE
If headers are off
PHASE
The second parameter has been set to phase angle.
Error
If the response message is longer than 300 bytes, a query error occurs.