5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
Note
The results of output when using the
*
WAI command differs from when not
using, since the :MEASure? query is not sequential command.
If the test frequency is set to 1 kHz:
When the
*
WAI command is not used
Transmission
:FREQuency 50;:MEASure?
The response message of this :MEASure? query is the measured value of
frequency at 1 kHz.
When using the
*
WAI command
Transmission
:
FREQuency 50;
*
WAI;:MEASure?
The response message of this :MEASure? query is the measured value of
frequency at 50 Hz.
Example
1. During normal testing
When querying the measured values for impedance (Z), phase angle (
θ
), parallel
capacitance (Cp), loss coefficient (D):
Transmission
:MEASure:ITEM 53,0;:MEASure?
Response
If headers are on
Z 31.981E+03,PHASE -88.05,CP 4.9736E-09,D 0.03405
If headers are off
31.981E+03,-88.05,4.9736E-9,0.03405
2. During comparator testing
When comparator testing for impedance (Z) and phase angle (
θ
).
Transmission
:PARameter1 Z;:PARameter3 PHASe
:COMParator ON
:MEASure?
Response
If headers are on
1,Z 31.981E+03,0,PHASE -88.05,-1
If headers are off
1,31.981E+03,0,-88.05,-1
The decision result of the first parameter is within limits, and that of the second
parameter is below the lower limit.
3. During scaling testing
When comparator testing for impedance (Z) and phase angle (
θ
).
Transmission
:PARameter1 Z;:PARameter3 PHASe
:SCALe ON
:MEASure?
Response
If headers are on
Z 31.981E+03,0,PHASE -88.05
If headers are off
31.981E+03,0,-88.05
Error
If the response message is longer than 300 bytes, a query error is generated.
During comparator testing and scaling testing, if parameters both first and second
are set to OFF, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.