28
_____________________________________________________________________________________________
5.3 Particular Commands
_____________________________________________________________________________________________
*
RST
Performs device initial setting.
Syntax
*
RST
Function
Resets the 3532-50. The items which are reset are listed below.
Test parameters Impedance (Z), phase angle (
θ
)
Test frequency 1 kHz
Test signal level Open circuit voltage mode (V mode)
V mode set value 1.00 V
CV (constant voltage) set value 1.00 V
CC (constant current) set value 10.00 mA
Limit function OFF
Voltage limit set value 5.00 V
Current limit set value 50.00 mA
Test range AUTO
Open circuit compensation OFF
Short circuit compensation OFF
Trigger setting Internal trigger
Trigger delay time 0 s
Averaging OFF
Test speed setting NORMAL
Beep sound setting ON for key input, OFF for comparator
DC bias function (3522-50 only) OFF
Cable length (3532-50 only) 0 m
Comparator
Comparator setting mode Both first and second parameters set to absolute
value
Absolute value set values
First parameter Upper and lower limit values: OFF
Second parameter Upper and lower limit values: OFF
Percent set values
First parameter Reference value: 1000
Upper and lower limit values: OFF
Second parameter Reference value: 10
Upper and lower limit values: OFF
Panel save All contents clear
Scaling Correction coefficient a: 1.0000, b: 0
Number of displayed digits 5 digits
Error
If the data parameters are set after this command, a command error occurs.