70
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
Range
number
Range (
Ω
)
1
0.1
2
1
3
10
4
100
5
1 k
6
10 k
7
100 k
8
1 M
9
10 M
10
100 M
:RANGe?
Queries the test range.
Syntax
:RANGe?
Function
Returns the test range setting as numerical value in NR1 format.
The numerical value corresponding to the test range and frequency which can be
set is as follows.
Example
Response
If headers are on
:RANGE 5
If headers are off
5
The test range has been set to range 5 (1 k
Ω
).
Error
If the response message is longer than 300 bytes, a query error is generated.