71
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
:RANGe:AUTO
Enables and disables the auto-range function.
:RANGe:AUTO?
Queries the autorange function enablement.
Syntax
:RANGe:AUTO
<data>
<data>
ON/OFF (character data)
Function
Switches between automatic and manual setting of test range.
ON Switches the automatic setting.
OFF Switches the manual setting.
Example
Transmission
:RANGe:AUTO ON
The test range is switched to automatic selection (auto-ranging).
Error
If <data> is other than character data described above, an execution error occurs.
Executing this command while the open or short circuit compensation is
performed generates an execution error.
Syntax
:RANGe:AUTO?
Function
Returns whether the test range is automatically set as character data.
ON, OFF
Examples
Response
If headers are on
:RANGE:AUTO ON
If headers are off
ON
Error