66
_____________________________________________________________________________________________
5.4 Commands Specific to the 3532-50, 3522-50
_____________________________________________________________________________________________
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Lp
Ls
D
Cp
Cs
θ
Y
Z
Measurement Register 0 (MR0)
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Unused Unused
B
X
Rp
G
Rs
Q
Measurement Register 1 (MR1)
:MEASure:ITEM?
Queries the test parameter for response of the :MEASure? query
during normal testing.
Syntax
:MEASure:ITEM?
Function
Returns the test parameter to response the :MEASure? query during normal testing
as bits <MR0> and <MR1>.
The items of two registers (MR0 and MR1) are as follows.
Example
Response
If headers are on
:MEASURE:ITEM 53,18
If headers are off
53,18
The test parameters to response have been set to impedance (Z), phase angle (
θ
),
parallel capacitance (Cp), loss coefficient (D), series resistance(Rs), reactance (X).
Error
If the response message is longer than 300 bytes, a query error is generated.