DPO6000
,
MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 93 -
19.Auxiliary Function Setting
19.1.Language
This oscilloscope supports menus in multiple languages as well as provides Chinese
and English help information, prompt messages and interface display.Press
Utility
Language,
rotate V0 to select the desired language, press V0 to select the language.
19.2.The switch of touch screen(not available for the machine without
touch screen function)
This series oscilloscope supports to open and close the touch screen manually,press
Utility
touch screen switch
to open and close the touch screen.
19.3.System Information
press
Utility
system information
to view the system information of your oscilloscope.
The system information includes the manufacturer, model, serial number as well as
software version and hardware version.
19.4.Pass/Fail
After selecting this type, in the pass/fail test, when the oscilloscope detects a waveform
that satisfied to the pass/fail rule and output mode, it will output a pulse through the
connector;When no waveform confirming to the above conditions is detected, the
connector no output.
19.5.Update the Firmware
DPO6000/MPO6000 can update the firmware by USB storage device,the whole process
is about 5 minutes.
Please update the oscilloscope by the following methods:
1.
Plug the USB flash disc in which a firmware program is saved in USB Host interface on
the front panel of the oscilloscope.
2.
Utility
Update
Update firmware
Manually select the firmware file
start the
upgrade process.
3.
Use V0 to select the .up file to update,and press V0 to select the updating file.
4.
Press F3 to start the updating process.
5.
The prompt message that says you have updated successfully will pop after updating.
6.
The oscilloscope will restart automatically.
19.6.Self-calibration
The self-calibration program can quickly make the oscilloscope reach the best working
state to get the most precise measurement values. You can perform self-calibration at any
time especially when the change of the environment temperature is up to or more than 5
℃
.
Make sure that the oscilloscope has been warmed up or operated for more than 30
minutes before the self-calibration. Disconnect all the input channel connections, and then
press
Utility
Self-Cal.
19.7.Function complex
Users can set the type of the signal output from the [Aux/Trigger Out] connector at the
rear panel.The function of trigger out and the result of pass/fail can also be realized
through "[pass/fail] → output to DDS" and “[signal generator] → burst → data source
selection through/fail or DSO Trigger”. For specific operation steps, please refer to the
introduction of "through failure" and "burst release" chapters.
Press
Utility
Function complex
to select the desired output type.
Trig Output
After this type is selected, the oscilloscope outputs a signal that can reflect the