DPO6000
,
MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 85 -
13.Burst
DPO6000/MPO6000 can output waveform with a specified number of cycles from a single
channel or two channels at the same time. When DPO6000/MPO6000 trigger,manual
trigger source control pulse string output.
1.Enable Burst Function
Press WaveGen
→
burst on the frontal panel
2.Set the Count Value
Press WaveGen
→
burst
→
count on the frontal panel
3.Set data Source
Press WaveGen
→
burst
→
data source on the frontal panel
Manual
:
At this time, "trigger" button is optional. Press "trigger" once and the signal source outputs
a waveform with a set number of cycles.
Pass/Fail:
Run the [pass/fail] test, open "output" and select "output to DDS". After the waveform is
output according to the set output mode, specify the waveform with the set number of
cycles output by the signal source.
DSO Trigger
:
Each time the waveform in an oscilloscope is triggered, the current signal source outputs
a waveform with a set number of cycles.