DPO6000
,
MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 59 -
waveform rapidly and intuitively,Simple measurement can be made by the division and
scale coefficient of the grid.
For example, a simple measurement can be made by calculating the relevant primary and
secondary scales and multiplying by the proportionality coefficient.
If calculated between the maximum and minimum of the waveform there are 6 main
vertical scale divisions, and given that the scaling coefficient is 50mV/ min, the peak-to-
peak voltage can be calculated as follows:
6 divisions×50mV/div=300mV
8.3.Auto Measurement
Quick Measurement after AUTO
When the oscilloscope is correctly connected and has detected valid input signal,press
AUTO
to set the waveform automatically and open the following function menu:
Single-period
: set the oscilloscope to display single-period signal automatically. Measure
the "Period" and "Frequency" of the current signal within a single period and display the
measurement results at the bottom of the screen.
Multi-period
: set the oscilloscope to display multi-period signal automatically. Measure
the "Period" and "Frequency" of the current signal within multiple periods and display the
measurement results at the bottom of the screen.
Auto measuring range
:set the ‘‘horizontal vertical scale,horizontal scale and vertical
scale’’
Data source
:set the auto channel to ‘‘only display’’ ,‘‘all’’.
8.4.One-key Measurement of 42 Parameters
Press the corresponding softkey under
MENU
at the left of the screen, to quickly measure
42 waveform parameters and realize "One-key" measurement. The measurement results
can be displayed at the bottom of the screen in various font sizes.
The time and voltage parameters icons in the measurement items and the measurement
results on the screen are always marked in the same color with the channel (
Measure
Source
) currently measured.
Note
: If the measurement result is displayed as "*****", it means that there is no signal
input in the current source or the measurement result is not within the valid range (too
large or too small).
The oscilloscope automatically executes all computing modes. This measurement uses
waveform recording points, which are more accurate than grid and cursor measurements.
The displayed readings of auto measurement are periodically updated with new data
obtained by the oscilloscope.
Press
Meas
to perform auto measurement. Automatic measurement of 42 parameters is
provided, and up to 4 parameters can be displayed at a time.
Perform the following steps,select voltage or time parameters for automatic
measurement:
1.Press
Meas
on the frontal panel to enter measurement function menu.
2.Press
signal source
softkey,and rotate V0 to select measuring channel.
3.Press
type
softkey, and rotate V0 to select the desired measuring parameters.
4.Press V0 to add measurement parameters, measurement parameters and
measurement values will be displayed in the directory, and the statistical status will be
updated in real time.
5.Turn off the statistics function,press statistics softkey and select turn off.
The measuring display area can display up to 4 measuring parameters and carry out
measurement according to the selection sequence.If the fifth measurement parameter is
added, the first measurement parameter will be deleted.
Note
: if the parameters do not match the measurement conditions, "****" will be displayed.
To Clear the Measurement parameters
Press clear all softkey to clear all the measurement parameters displayed on the screen.
Statistics function
Count and display the current value, average value, minimum value, maximum value and