MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 102 -
Technical Specifications
All specifications apply to the DPO6000/MPO6000 series oscilloscopes,to see the details
in the last part of this chapter. Before checking an oscilloscope see if it complies with
these specifications, make sure it meets the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Self Calibration operation must be performed through the Utility menu if the
operating temperature changes by more than 5
℃
.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted “typical”.
Oscilloscope Specifications
Acquisition
Acquisition Methods
Real-time sample
Real-time
sample
rate
1 GSa/s(single channel) 500 MSa/s(two channels)
250 MSa/s(three/four channels)
;
Peak detection
Analog channel 4ns
Note
:digital channels don’t support
Average mode
Analog channel
All channels reach N time samplesat the same time,N can be selected
from 2
、
4
、
8
、
16
、
32
、
64
、
128
、
256
、
512 and 1024.
Note
:digital channels don’t support
High resolution
Up to 12bit
Note
:digital channels don’t support
Minimum test pulse
width
8ns
Memory depth
Single channel 64M/128M[EDU]
Two channels 32M/64M[EDU]
Three,four channels 16M/32M[EDU]
Five,six,seven,eight channels 8M/16M[EDU]
Input