DPO6000
,
MPO6000 Series Digital Phosphor Oscilloscope Instruction V1.1
- 60 -
root mean square error of up to 4 open measurement items.
1. Press
Meas
on the frontal panel to enter measurement function menu.
2.press
statistic
softkey and select open.
All Measurement
All measurement could measure all the parameters of the current signal source and
display the results on the screen.
Follow these steps to measure all parameters:
1.Press Meas
on the frontal panel to enter measurement function menu.
2.press
all measurement
softkey and select open.
3.press
signal source
softkey to select the measurement signal source(CH1~CH4).
8.5.Cursor Measurement
Cursors:
This method allows the user to make measurements by moving the cursor.
Cursors always appear in pairs, and the readings displayed are the measured values.
There are two types of cursors: amplitude cursor and time cursor.
Amplitude cursor: the amplitude cursor is displayed as a horizontal dotted line to
measure the parameters in the vertical direction.
Time cursor: the time cursor is displayed as a vertical dotted line to measure the
parameters in the horizontal direction.
The cursor measurement includes two modes
:
Manual mode and Tracking mode.
1.Manual mode
:
Horizontal cursors or vertical cursors appear in pair to measure time or voltage, and the
position of the cursors can be manually regulated. The signal source should be set as a
waveform to be measured before the cursors are used.
2.Tracking mode
:
A horizontal cursor is intersected with a vertical cursor to form a cross cursor. The cross
cursor is automatically located on the waveform, and the horizontal position of the cross
cursor on the waveform is regulated by selecting “Source A” or “Source B” and rotating the
V0 knob. The coordinates of the cursor point will be displayed on the screen of the
oscilloscope.
Press
【
CURSOR
】
to enter the cursor menu
Options
Settings
Annotation
Mode
manual
tracking
Select a measuring cursor and display.
Data Source
CH1~CH4
MATH
Select a signal source and measure by cursor.
Select cursor
AX(BX)
AXBX
AY(BY)
AYBY
The selected highlight cursor can move freely.You can
select 2 cursors and move them at the same time.The
frame behind the cursor display the position of the cursor.
Move the cursors
:Press the key near the select cursor and rotate V0 to move the
cursor.The cursor can only be moved when the cursor menu is open.
8.6.Digital Voltage Meter(DVM)
DVM is mainly used for voltage and frequency measurement,and monitoring the signal
when the oscilloscope is running or stopping.
DVM has 5 digits AC RMS
、
DC
、
AC+DC RMS voltage measurement and 6 digits
frequency measurement.
8.6.1.Enable the digital voltage meter
Press Meas
→
DVM
→
enable
【
on,off
】
on the frontal panel.
8.6.2.Set the data source
Press Meas
→
DVM
→
data source
【
CH1,CH2,CH3,CH4
】
on the frontal panel.
8.6.3.Set types