99
CONTROL MODES
Pr
op
or
tiona
l C
ont
rol I
nt
er
ac
tion w
ith C
ont
rol
ler S
ta
tu
s
Function /
Mode
Hold
Cleaning
Edit
Calibration
Manual
Error
Measure
Start up
Hold
Alarm
Control Output
o to 100% On
from Control
Period
0%
0%
0%
0%
0%
0%
0%
0%
Relay Assigned
to Set P
oint
Control Output
On for the
time control
output is On
Off
Off
Off
Off
Off
Off
On or Off
Off
Analog Output
Assigned to
Set P
oint
Control Output
Scaled value
of control
output
Scaled value
of control
output
Scaled value
of last control
output or a
fixed value of
control output
Scaled value
of control
output or 22
mA
, if option
configured
Scaled value
of last control
output or a
fixed value of
control output
Scaled value
of last control
output or a
fixed value of
control output
Scaled value
of last control
output or a
fixed value of
control output
Any value in
range 0 to
22 mA
Scaled value
of control
output