HI510 EVENTS MANAGEMENT SYSTEM
Event & Log Event Code
Description
Hold Mode
Logging
Analog Output (AO) Behavior
Status LED
WARNING_HIGH_SECONDARY
0x00000800
Generated
when
CH secondary parameter
is over
CH
secondary parameter AL
ARM High set value but the
mask time has not expired to became alarm
NO
YES
As it is configured
WARNING_LOW_SECONDARY
0x00001000
Generated when CH secondary parameter is below
CH secondary parameter AL
ARM High set value but
the mask time has not expired to became alarm
NO
YES
WARNING_CTRL_CAL_
CHANGED
0x00002000
Displayed after Calibration update
NO
YES
WARNING_CTRL_CH1_SP1_
SETTINGS_INCOMPLETE
0x00004000
Generated if Set point1 is enabled but no output is
assigned
NO
YES
WARNING_CTRL_CH1_SP2_
SETTINGS_INCOMPLETE
0x00008000
Generated if Set point2 is enabled but no ouput is
assigned
NO
YES
WARNING_RTC_SET_TO_
FIRST_VALUE
Log event code: 0x00010000
Generated when RTC was initialized
NO
YES