53
MENU
11.1.4. Alarm Settings
This menu is used to define the operating limits of the process. The setting thresholds configured in
this submenu control the Alarm relay. If Alarm becomes active, control stops. Both pH (or ORP) and
Temperature are configured in this submenu.
Navigation
• Press MENU key ( ) from the Measure mode.
• Press
Setup
from Channel parameter.
• Press
Setup
with Alarm Settings highlighted, and the alarm submenu will open.
• Press the
keys to move between options.
• Press the back key ( ) to return to the menu without saving.
Note:
It may be necessary throughout the Setup:
at prompt, enter the passcode
at prompt, press
YES,
to place unit in HOLD.
Note:
We suggest users make configuration changes from the beginning of the menu structure
going forward, because the menu references parameters that were set earlier in the submenu.
When completed, return to the “other” parameter and set that up also. Alarm can be configured
for both measurement and Temperature.
Parameter (Probe Specific)
Option:
pH (ORP), Temperature
With Parameter selected, press the corresponding virtual key to toggle between the options.
Note:
The second Parameter can be set up also.
Alarm High Enabled
Option:
Enabled, Disabled
With Alarm High Enabled selected, press the corresponding virtual key to toggle between enable or
disable options.