![Hanna Instruments HI510 Instruction Manual Download Page 84](http://html1.mh-extra.com/html/hanna-instruments/hi510/hi510_instruction-manual_3772999084.webp)
84
MENU
Remote Control
Option:
Enabled, Disabled
This option allows the user to Enable Remote Control. This must be enabled if using the PC application
HI92500
.
With Remote Control selected, press the corresponding virtual key, to toggle between options. The
check mark confirms the enabled parameter.
RS485 Address
Option:
01 to 99
This option allows the user to set the RS485 Address. The controller and the PC application
HI92500
must have the same RS485 Address to communicate.
With RS485 Address selected, press
Set,
to modify. Press the key to enter the digit. Keep the
key (or key) pressed to increment (or decrement) by one, every second. Press
CFM,
to save.
RS485 Baud Rate
Option:
9600, 19200, 38400, 57600, 115200, 256000
This option allows the user to set the desired speed for the serial communication (baud rate) in bps.
The controller and the PC application
HI92500
must have the same baud rate.
With RS485 Baud Rate selected, press
Modify,
for the drop-down list to display. Press the
keys
to navigate between options. Press
Select
, to save.
Startup Delay
Option:
1 to 30 minutes
Startup Delay is a timer used to prevent Control functions (Relays and pH (ORP) and Temperature
configured Analog Outputs) from functioning during controller startup.