121
HI510 EVENTS MANAGEMENT SYSTEM
Event & Log Event Code
Description
Hold Mode
Logging
Analog Output (AO) Behavior
Status LED
WARNING_FUSB301_FAIL
0x00020000
WARNING USB Controller FAIL
NO
YES
As it is configured
WARNING_CTRL_SET_
CHANGED
0x00040000
Control settings changed
NO
YES
WARNING_LOG_INTERVAL_
CHANGED
0x00100000
Generated when the Lot logging time-interval has
changed
NO
YES
WARNING_LOT_LOG_FULL
0x00800000
Generated when current file has reached 8600
records. A new lot log will be generated.
NO
NO
WARNING_LOT_LOG_MA
X_
INDEX_ASSIGNED
0x01000000
Generated when the Lot logging max. file
index (99) was reached (the next lot file will
automatically delete the oldest one)
NO