56
MENU
4. The state next to the Menu item will change to
Man On
(or
OFF
)
5. Press the key, to exit the parameter
When in Manual Hold:
• Hold is displayed in the Title & Status area
• The primary measurement value is displayed blinking
• The HOLD LED is on
• Any relay configured for Hold; relay LED will be on with associated relay enabled
• All the alarm signals (LED, alarm relay) are suspended
• Analog Outputs will be at configured state (see Analogs)
11.2.2. Configuring External Hold Trigger
Navigation
• From Menu, use
keys to select Hold Mode Man. On.
• With Hold Mode Man. On selected, press
Setup
, to enter the screen.
Hold Input Enabled
Option:
Enabled, Disabled
With function selected, press
Enable
or
Disable
, to toggle between the two options. The check mark
confirms Hold Input enabled.
Hold Input
Hold mode can be triggered using external trigger inputs. This is a read-only parameter that indicates
what Inputs are configured to initiate Hold mode. If an input is selected, the selected input is displayed.
To change the input assignment for Hold Input, return to the top level Menu structure and select
Inputs. To return to the menu without changing, press the key.