118
HI510 EVENTS MANAGEMENT SYSTEM
Event & Log Event Code
Description
Logging
Stop
Ctrl.
Analog Output (AO)
Alarm Relay &
LED
Status LED
AL
ARM_UNDER_RANGE_
PRIMARY
0x00000800
Generated during measurement cycle when probe main reading is in under range status
YES
YES
AO assigned to Ctrl.SetP
oint – scaled
value of Ctrl.SetP
oint output
AO assigned to parameter – scaled value of parameter or 22mA
, if this option is
enabled
On
AL
ARM_OVER_RANGE_
SECONDARY (T)
0x00001000
Generated during measurem. cycle when probe temperature reading is in over range status
YES
probe use manual T
.
–
AL
ARM_UNDER_RANGE_
SECONDARY (T)
0x00002000
Generated during measurem. cycle when probe temperature reading is in under range status
YES
probe use manual T
.
–
AL
ARM_HOLD_IN
0x00100000
Generated each time HOLD input condition is present (alarm hold)
YES
YES
On
Generated manual (silent hold)
YES
YES
Generated by Menu /User cal. (silent hold)
YES
YES
AL
ARM_OVERTIME
Set point 1: 0x00000010 Set point 2: 0x00000020
Generated each time overtime set value is exceeded with 100%
YES
YES
v