![Hanna Instruments HI510 Instruction Manual Download Page 122](http://html1.mh-extra.com/html/hanna-instruments/hi510/hi510_instruction-manual_3772999122.webp)
122
HI510 EVENTS MANAGEMENT SYSTEM
ER
RO
RS
Event & Log Event Code
Description
Hold
Mode
Logging
Stop
Ctrl.
Analog Output (AO) Behavior
Alarm Relay & LED
Status LED
0-20 mA
4-20 mA
ERROR_PROBE_NO_FCAL
0x00020000
Probe is not factory calibrated or
factory calibration is corrupt
YES
Event
YES
0
4
ERROR_5V_POWER
0x00000200
Generated when and reading is not
in range
YES
Event
YES
0
4
ERROR_MULTIPLE_POWER
0x00000400
0x00000100
0x00000800
0x00001000
or the sum of them
Generated when there is an error
on RS-485 power supply, 3V power
supply, probe 1 power supply, or
probe 2 power supply is not in range
YES
Event
YES
0
4
ERROR_AO_POWER
0x00004000
Generated when Analog Outputs (AO)
power supply is not in limits
YES
Event
YES
0
4
ERROR_IO_POWER
0x00010000
Generated when smart Input power
supply is not in limits
YES
Event
YES
0
4