MENU
Setup for Proportional control
• For the drop-down list to be displayed, press
Setup
.
• Press the
keys to move between Mode, Deviation, Control Period and Dead Band.
• Select Mode and press the virtual keys to choose Prop. Low or Prop. High.
• Press the key to select Deviation.
• With Deviation (default value 1.00 pH) highlighted, press
Set
. The present value will blink
permitting editing, within minimum 0.02 pH and maximum 9.00 pH, using the
keys.
• Press
CFM,
to save.
• Press the key, to select Control Period.
• With Control Period (default value 1.00 minute) highlighted, press
Set
. The present value will blink
permitting editing, within minimum 10 s and maximum 30.00 minutes, using the
keys.
• Press the key, to select Dead Band.