51
MENU
• Press
CFM,
to save
• Press the key, to select
Reset Time
.
• With Reset Time (default 16:40 h) highlighted, press
Set
. The present value will blink permitting
editing, within minimum 1 minute and maximum 16:40 h, using the
keys. The default
value disables the Integrative contribution.
• Press the key, to select
Rate Time
.
• With Rate Time (default value 0 s) highlighted, press
Set
. The present value will blink permitting
editing, within minimum 0 s and maximum 16:40 h, using the
keys. The default value
disables the Derivative contribution.
• Press
CFM,
to save.
• Press the key, to select Dead Band.
• With Dead Band (default 0.20 pH) highlighted, press
Set
. The present value will blink permitting
editing, within minimum 0.00 pH and maximum Deviation value divided by 5, using the
keys.
• Press
CFM,
to save.
• Press the key, to select Dead Band Gain.
• With Dead Band Gain (default 0%) highlighted, press
Set
. The present value will blink permitting
editing, within minimum 0% and maximum 100%, using the
keys.
• Press
CFM,
to save.