50
MENU
• With Dead Band (default value 0.02 pH) highlighted, press
Set
. The present value will blink
permitting editing, within minimum 0.00 pH and maximum Deviation value divided by 5, using
the
keys.
• Press
CFM,
to save
• Press the key, to exit Setup.
Setup for PID control
• For the drop-down list to be displayed, press
Setup
.
• Press the
keys to move between Mode, Deviation, Control Period, Reset Time, Rate Time,
Dead Band and Dead Band Gain.
• Mode: press the virtual keys, to choose PID Low or PID High.
• Press the key, to select Deviation.
• With Deviation highlighted, press
Set.
The present value will blink permitting editing using the
keys.
• Press
CFM,
to save.
• Press the key, to select Control Period.
• With Control Period highlighted, press
Set
. The present value will blink permitting editing using
the
keys