EXTERNAL CONTROL
173
External Resistance Control - Overview
Background
External resistance control of the CC, CR, CV and
CP modes is accomplished using the J1 connector
on the rear panel.
A resistance of 0kΩ-10kΩ is used to control the
input current, voltage, resistance or power on the
PEL-3000(H).
The input can be configured to vary in proportion
to the external resistance or the inverse. See page
174 for more details on proportional and inverse
resistance control.
Note
Exceeding 11.8k
Ω
will cause an EXT.OV alarm
message which will reset the voltage output to 0 until
the external resistance is reduced back down below
11.8k
Ω
.
Connection
When connecting the external resistance source to
the J1 connector, use a ferrite core and use twisted
pair wiring.
PEL-3000(H)
EXT-R
J1
connector
3
1
Input
Terminals
Ferrite Core and
twisted wiring
Pin1 → EXT-R
Pin3 → EXT-R
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...