FUNCTION MENU
137
Description
Fast Sequence configuration is split into Timing
Edit configuration and Data Edit configuration.
Timing Edit configuration is used to configure all
the settings that are common to all the steps of the
fast sequence. This includes settings such as the
mode, range, loops and time base.
Data Edit configuration is used to create the actual
steps used in each sequence.
See below for a description of each.
Timing Edit
Overview
A Fast Sequence contains the following timing
settings for each sequence:
Setting
Setting Range
Description
Memo
12 characters
A user-created note for the
currently selected sequence.
Mode
CC, CR
Operating mode for the
sequence.
Range
ILVL
Low I range, low V range
IMVL
Middle I range, low V range
IHVL
High I range, low V range
ILVH
Low I range, high V range
IMVH
Middle I range, high V range
IHVH
High I range, high V range
Loop
Infinity, 01 - 9999
Sets the amount of times to loop
the selected sequence.
Last Load
OFF, ON
Set the load condition after the
end of the sequence.
Last
0.000000
The load setting for when Last
Load is set to ON.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...