PEL-3000(H) User Manual
224
CR Mode
Load Input
Voltage
Load Current
CR Value
CR+CV Mode
When CR+CV mode is enabled, the unit will act as
constant resistive load after the input voltage is
greater than the user-defined CV level. At the CV
level, the unit works as a constant voltage load.
This mode effectively creates a voltage ceiling
before the unit operates in CR mode. The diagram
below illustrates this.
CR + CV Mode
Load Input
Voltage
Load Current
CV level
CR Value
Note that when the source voltage is less than the
CV level, no current will flow due to a very high
impedance.