FUNCTION MENU
121
4.
Select
Memory
and select which memory
location to load for the selected step.
Settings loaded from the memory location will
be used for the selected step.
The same memory location can be used for
multiple steps.
Memory
M001 - M256
5.
Set the
Run
setting for the step.
By default RUN is set to Skip.
The Auto setting will automatically start and go
onto the next step.
The Manual setting will wait for the user to
press
Next[F2]
before running the step.
Run
Skip, Auto, Manual
6.
Choose the
On-Time
in seconds.
The on-time setting determines how long the
load is turned on for the selected step.
The on-time is defined as the total test time
minus the off-time.
On-Time
0.1 - 60 seconds
7.
Choose the
Off-Time
in seconds.
The off-time setting determines how long the
load is turned off between the end of the
current step and the start of the next step.
The off-time is defined as the total test time
minus the on-time.
Off-Time
Off, 0.1 - 60 seconds
8.
Choose the
P/F-Time
(pass/fail time) in
seconds.
The P/F-Time refers to the P/F delay time. This
delay time includes the 0.06 P/F start test time,
as shown in the timing diagram on page 119.
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...