PEL-3000(H) User Manual
152
Parameters
OPP. No
Selects one of 12 OPP test
setup memories.
Memo
A user-created note for the
currently selected OPP
function.
Range
High(CP Mode High)
Middle(CP Mode Middle)
Low(CP Mode Low)
Start Watt (Start W)
Starting watt value for the
test.
End Watt(End W)
The watt value that will end
the test. The value must be
higher than the OPP value
of the DUT you are testing.
This parameter is used as a
fail-safe for if the over
power protection of the
DUT fails. If the measured
watt is reaches End Watt
value it would then indicate
that the power supply OPP
failed.
Step Watt(Step W)
Sets the step resolution of
the watt.
Step Time(Step T)
Sets the execution time of
each step. (10ms to 50s)
Trig Delay
Time(Delay)
Sets a delay corresponding
to the time a Trig Voltage
can be expected after each
step Watt is applied (the
delay time must be less
than the Step time).
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...