PEL-3000(H) User Manual
68
Operation
1.
Make sure the load is off.
2.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Dyna. Level
setting.
Range:
Value, Percent
Display:
Percent Setting
Mode
CP
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
Timer1
0.025
Level
50.0
Set
1.00
LOAD
RS232
Fine
01/Oct/2018
ms
%
W
Set level
% from the Set value
Example
Level
Time
Set
Level
Timer1 Timer2 Timer1
Dynamic Mode:
Dyna. Level = Percent
Select the Switching Time Configuration for Dynamic Mode
Description
The switching time for dynamic mode can be
configured to switch between two preset on-times
(Timer1, Timer2) or by setting a switching
frequency and duty cycle.
Operation
1.
Press
Main
>
Configure[F5]
>
Other[F2]
and
set the
Dyna. Time
setting.
Range:
T1/T2, Freq. Duty
Summary of Contents for PEL-3000 Series
Page 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Page 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Page 41: ...GETTING STARTED 39 Fasteners 2 ...
Page 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Page 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Page 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Page 231: ...APPENDIX 229 PEL 3111 ...
Page 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Page 233: ...APPENDIX 231 PEL 3041H ...
Page 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Page 235: ...APPENDIX 233 ...